Klietz, Marie-Luise (2022): Einfluss der Lagerungsdauer und Lagerungsmethodik von Testserum aus dem Kniegelenksimulator auf die Partikelanalyse. Dissertation, LMU München: Medizinische Fakultät |
Vorschau |
PDF
Klietz_Marie-Luise.pdf 20MB |
Abstract
Introduction: While complication rate of knee prosthesis is strongly associated with their particle wear, no attention has been paid on the storage method and time up to now. Material and Methods: We therefore collected bovine serum after 500.000 cycles in the knee joint simulator and evaluated particle characteristics for a period of one year at different storage methods. Results: Our study showed that the storage time of the test serum has significant influence on the characteristics of the particle wear, while storage method does not influence particle characteristics. Conclusion: The evaluation time of the particle serum should be standardized to ensure comparability of future studies.
Dokumententyp: | Dissertationen (Dissertation, LMU München) |
---|---|
Themengebiete: | 600 Technik, Medizin, angewandte Wissenschaften
600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin und Gesundheit |
Fakultäten: | Medizinische Fakultät |
Sprache der Hochschulschrift: | Deutsch |
Datum der mündlichen Prüfung: | 11. April 2022 |
1. Berichterstatter:in: | Wegener, Bernd |
MD5 Prüfsumme der PDF-Datei: | bb5e33d2171fbd3f280b93a0194f8b2a |
Signatur der gedruckten Ausgabe: | 0700/UMD 20362 |
ID Code: | 29788 |
Eingestellt am: | 12. May 2022 09:06 |
Letzte Änderungen: | 12. May 2022 09:06 |